GE Healthcare · MRExcite2245100741345574386
ACGD Gradient Driver Test Fails Due to the SGA ACGD Power Supply.
Symptom
ACGD Gradient Driver Static Test Fails ACGD Gradient Driver Test Fails
Error message
2245100 GRADIENT DRIVER TESTS - Static Test Failed on % Axis Fri Jun 11 17:57:56 2004 Error: 2245100 Host: lx-mr Process: diagCoord File: grad_lib.c Line: 416 Diagnostic: GDT Stat&Dyn Iteration: 1 of 1 Action: 1 Task ID: 0xfc9990 Time: 741345 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on X A Fri Jun 2 17:37:01 2006 Error: 2245100 Host: lxmr3 Process: diagCoord File: grad_lib.c Line: 921 Diagnostic: GDT Static Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c6c8 Time: 574386 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Test Failed on the Z Axis. VBUS_700 had 2 failures out of 39. Ready Static Mode-> this happened on the z axis- tech swapped the power inputs and the error moved-ordred SGA power supply
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.