GE Healthcare · MRExcite2252720
MDS Link failure due to Power Supply 2 in MGD chassis.
Symptom
Scan Aborts with MDS Link Error
Error message
2252720 Sequence playout error. The SRF signaled a SRF power fault error. see long description Error: 2252720 Host: lx-mr Process: NSP File: AGP:agp drv.c Line: 510 Sequence playout error. The SRF signaled a SRF power fault error.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2294300-8 MGD Power Supply ***OBSOLETE PART Case # 11430551*** See Long Description ***OBSOLETE PART Case # 11430551*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-254 Suitable alternative part is 5181495-3. Please see latest appropriate FRU manual for product in question. 212002950 February 20, 2009.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.