PS2 failure in MGD chassis, dragging down the 3.3 & 5 volt power.

Symptom

Scan Stops With SRF Signaled a TRF Power Fault Error.

Error message

2252721 - The SRF signaled a TRF power fault error. Error: 2252721 Host: lx-mr Process: NSP File: AGP:agp drv.c Line: 519 Sequence playout error. The SRF signaled a TRF power fault error.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2294300-8 MGD Power Supply ***OBSOLETE PART Case # 11430551*** See Long Description ***OBSOLETE PART Case # 11430551*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-254 Suitable alternative part is 5181495-3. Please see latest appropriate FRU manual for product in question. 212002950 February 20, 2009.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.