Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova3602074

The DL Digital Leader console is locked, the system will not allow an End exam.

Symptom

The DL Digital Leader console is locked, the system will not allow an End exam. During system use the Digital leader console is locked and it will not allow the customer to end an exam. The DL is displaying the message "acquisition in progress" and the pop up will not allow user input.

Error message

3602074 DO XRTechMgr Jedi: Prep too long Debug: jedi-error-code=704h JEDI Error Code: 0704H Heater / Rotor hold too long Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: READY_FOR_EXPOSURE 3602074 DO XRTechMgr Jedi: Prep too long Debug: jedi-error-code=704h JEDI Error Code: 0704H Heater / Rotor hold too long Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: READY_FOR_EXPOSURE This warning occurs when the operator holds the prep too long without exposure.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100