Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova1200109

Innova or Health Alert message detector " Coolant Fluid Low Level", Detector Conditioner Tank Evaporation or c…

Innova or Health Alert message "Coolant Fluid Low Level", Conditioner errors - Low fluid level

Symptom

Innova or Health Alert message detector " Coolant Fluid Low Level", Detector Conditioner Tank Evaporation or conditioner connection leaking Innova or Health Alert message detector "Coolant Fluid Low Level", Detector Conditioner Tank Evaporation or conditioner connection leaking. Replace with kit 2379172. Try again or press SYSTEM RESET. This is no longer an error message on Unity SP3.x+, Just low level warning logged in the error log. Innova or Health Alert message "Coolant Fluid Low Level", Low fluid, Detector Conditioner Tank Evaporation or conditioner connection leaking. Replace with kit 2379172 This is often indicated by an Innova Promise alert of low conditioner fluid. 212009582 01/14/2010 20:59 GMT Innova or Health Alert message "Coolant Fluid Low Level", Conditioner errors - Low fluid level This is no longer an error message on Unity SP3.x+, Just low level warning logged in the error log.

Error message

error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH, major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service" error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH , major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service"

INX100: 1200109, Class = WR error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = WR, major Function /= XRImDet System impact: None. Displayed message : "Detector chiller fluid level is low. Call Service.". 1200109, Class = WR error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = WR , major Function /= XRImDet System impact: None. Displayed message : "Detector chiller fluid level is low. Call Service.".

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
X-RAY DETECTOR COOLANT X-RAY DETECTOR COOLANT 2369908 FLS - X-RAY DETECTOR COOLANT CONTAINS 1 32 OUNCE BOTTLE APOLLO DETECTOR COOLANT - SINGLE LITER BOTTLE PROPYLENE GLYCOL LABELED 2352900 FOR USE IN ALL STATIONARY APOLLO DETECTORS INCLUDING SENO 2000D X-RAY DETECTOR COOLANT 2369908 FLS - X-RAY DETECTOR COOLANT CONTAINS 1 32 OUNCE BOTTLE APOLLO DETECTOR COOLANT - SINGLE LITER BOTTLE PROPYLENE GLYCOL LABELED 2352900 FOR USE IN ALL STATIONARY APOLLO DETECTORS INCLUDING SENO 2000D
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues