Error CANNOT SWITCH RECEPTORS DURING AN ACQUISITION on monitors

Symptom

Error CANNOT SWITCH RECEPTORS DURING AN ACQUISITION on monitors

Error message

50122 POS : FLUORO: FOOT OR HANDLE PREP SWITCH PERMANENT VALID HWStatus=0000 User=0000 0000 4AC7 50122 : Recommended Actions: Root Cause : Component(s) - Footswitch Handswitch. An error has been found on Footswitch or Handswitch on Positioner software. Recommended Actions: Run RTAC SIB RTBus diagnostics.

70082 CANNOT SWITCH RECEPTORS DURING AN ACQUISITION HWStatus=0000 User=0000 0000 6700 70082 : Recommended Actions: Root Cause : Software problem. This error is due to parameter out of range, bad state, bad pointer detected in the Titan application software. On the Advantx an attempt was made to switch receptors at the console or while fluoro or record exposures are in progress. Recommended Actions: If an attempt was made to switch receptors at the console or while fluoro or record exposures are in progress wait until exposures are complete before attempting the switch again. If recovery class of error is IH or AR , then reload Titan subsystem software. If problem persists, then replace Titan. If recovery class is DO, then it could be only an effect of other problem.

Applies to

Models
Innova
Area
X-ray Vascular