GE Healthcare · X-ray VascularInnova22000102210001
3D acquisition, x-ray aborts at start of spin.
Symptom
3D acquisition, x-ray aborts at start of spin.
Error message
2200010, error Text = RT changed Mask signal problem followed within 2 seconds by 2210001, error Text = Generic SW error for positioner subsystem System impact: Bolus or 3D acquisition aborted. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service" 2200010, error Text = RT changed Mask signal problem, debug Text += did not receive RTB_TABLE_MOVE_STATUS_HIGH since startPrepare, recovery Class = AR, major Function /= XRPatPos followed within 2 seconds by 2210001, error Text = Generic SW error for positioner subsystem, debug Text += RTLine is probably stuck, recovery Class = AR, major Function /= XRPatPos System impact: Bolus or 3D acquisition aborted. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service" For Bolus X-ray abort after the first 1 or 2 Mask exposures. For 3D X-ray abort as soon as acquisition is started. error Code = 2200010, error Text = RT changed Mask signal problem, debug Text += did not receive RTB_TABLE_MOVE_STATUS_HIGH since startPrepare, recovery Class = AR, major Function /= XRPatPos followed within 2 seconds by error Code = 2210001, error Text = Generic SW error for positioner subsystem, debug Text +=
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.