GE Healthcare · X-ray VascularInnova10500951050002

All Gantry axis's will not drive

Symptom

INX100, Lost All Gantry axis's drive or NO movements The Gantry axis's will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor The Gantry axis' will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor

Error message

1050095 Pivot AXIS CONTROL LOOP Failsafe error 1050095 DO Pivot AXIS CONTROL LOOP GantryAxis:null Pivot_AXIS_CONTROL_LOOP Failsafe ERROR 1050095 DO Pivot AXIS CONTROL LOOP GantryAxis null Pivot_AXIS_CONTROL_LOOP Failsafe ERROR Pivot: axis control loop problem: A problem has been detected by the Pivot MCB board in its control Loop

1050002 DO GantryAxis BUDDY Larm AXIS ENABLE disable from buddy 1050002 DO BUDDY Larm AXIS ENABLE GantryAxis null BUDDY_Larm_AXIS_ENABLE disableFromBuddyN This error is caused by the Buddy of the LARM (Pivot MCB) disabling the enable of the LARM, to prevent Gantry motion during a failure. The disable is issued when the Buddy AMP detects a failure on the LARM or the Buddy (Pivot) has failed 1050002 DO BUDDY Larm AXIS ENABLE GantryAxis null BUDDY_Larm_AXIS_ENABLE disableFromBuddyN This error is caused by the Buddy of the LARM (Pivot MCB) disabling the enable of the LARM, to prevent Gantry motion during a failure. The disable is issued when the Buddy AMP detects a failure on the LARM or the Buddy (Pivot) has failed.

Applies to

Models
Innova
Product
X Innova 2100 Do not use Innova 2100iq System Innova 2100
Software
Unity 3D M4 SP2 Unity 3D M4 SP2 Fwd Prod. DL Release : DL_UNITY_4.17.5 Ghost DL : 21.22.1 DL : DL_Unity_4.17.5 DL Atlas Core : 4.2-8_5 RTAC Atlas Core : 4.2-8_5 IDC : 5.26 Calypso : 4.37 Calypso Uptime : CLYUPT.UNITY_4.15 Positioner : RR POS LBY 1.4.4U_13 Jedi : P636RC1
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.