Either the Ethernet card in the HP Host PC computer or the entire computer was replaced.
Symptom
How to obtain new Scan Option Keys in the event the Ethernet card is replaced. All scan option keys are invalid because the FE either replaced the Ethernet card in PCI slot 5 of the HP Host PC computer, or else the FE replaced the computer, and failed to swap the old Ethernet card into the new computer upon replacement. The "Host ID" obtained electronically from the Ethernet card is used to encode the scan option keys. With a new card there is a mismatch and all previously encoded keys will not work. How should the FE get new keys if this happens? All scan option keys are invalid because the FE either replaced the Ethernet card in PCI slot 5 of the HP Host PC computer, or else the FE replaced the computer, and failed to swap the old Ethernet card into the new computer upon replacement. The "Host ID" obtained electronically from the Ethernet card is used to encode the scan option keys. With a new card there is a mismatch and all previously encoded keys will not work. How should the FE get new keys if this happens?
Error message
eLicense Invalid eLicense Invalid eLicense Invalid
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.