The Omega table elevation will not drive down vertically.

Symptom

The Omega table elevation will not drive down vertically. When the customer selects the down elevator motion, the enable works, but the table does not lower. The elevator up function is normal. The Omega table top will not drive down vertically. When the customer selects the down elevator motion, the enable works, but the table does not lower. The elevator up function is normal.

Error message

1050233 POS BOOT / Debug - One MCB extended PRD failed

1050259 Stepper Board Initialization Error 1050259 STEPPER INIT BOARD ERROR WR GantryAxis 0 STEPPER_INIT_BOARD_FAILURE 1 0 - -- OMEGA4/5 -- motorization licensed asked by System, BUT stepper not available Z:/heroic/table_geometry_task.c 480 1050259 STEPPER INIT BOARD ERROR WR GantryAxis 0 STEPPER_INIT_BOARD_FAILURE 1 0 - -- OMEGA4/5 -- motorization licensed asked by System, BUT stepper not available Z:/heroic/table_geometry_task.c

1050278 CPU Stepper Board Serial Communication Failure 1050278 CPU STEPPER BOARD SERIAL COM DO GantryAxis Reset From Back Plane Component(s)_Positioner_CPU_Table_Serial_communication 1 0 - Stepping board PRD no response Z:/heroic/table_geometry_task.c 1168

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Omega IV Cardiac table Omega IV Cardiac table Omega IV cardiac table Omega V Table Omega V Table Omega V Table Innova 2100–IQ Case Innova 2100–IQ System can be supplied with an OMEGA IV Compact patient table, an OMEGA V long patient table, which includes a table side control with Contour Filter, or a InnovaIQ patient Table. Innova 3100/3100-IQ and 4100/4100-IQ Case Innova 3100/4100 Systems can be supplied with an OMEGA V Long patient table or InnovaIQ patient Table, which includes a table side control with Contour Filter. Illustration 2-1: Omega patient table (with TSSC)
Area
X-ray Vascular

Full cause & step-by-step resolution