No fluoro exposure, no aquisition exposure during a case.

Symptom

No Acquisition exposure, fluoro or record will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Innova 2100iq system will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Error 3200011

Error message

50125 hw-status=00AB SH master - Joystick motion: TABLE DRIVING (longitudinal & up/down) error= 4ACA, hw-status=00AB, Error Code = 50125 , hw-status=00AB ,error Text = POS TSUI: SMB or SMH error, debug Text += pos-error=4ACA, hw-status=00AB, recovery Class = DO, major Function /= LC Gantry SH master - Joystick motion: TABLE DRIVING (longitudinal & up/down)

ADV/LC; 7 IH POS SYSTEM ERROR: NOT COMMUNICATING WITH NODE Debug: fct-status=21 hw-status=0 pos-error=4020 usr-status=0

70007 "CANNOT DO FLUORO, POSITIONER IS NOT AVAILABLE

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Area
X-ray Vascular

Full cause & step-by-step resolution