Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova600200

collimator 24v supply measures 47 volts

Symptom

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset. Customer needs to boot again to recover. When Innova system has a actor issue.

Error message

600200 Collimator This error is due to a time out on Communication to Collimator. 600200: Root Cause: Component(s) - Collimator This error is due to a time out on Communication to Collimator. Recommended Action: Make sure that the positioner has booted correctly. Run RTAC subsystem//P 600200: Root Cause: Component(s) - Collimator This error is due to a time out on Communication to Collimator. Recommended Action: Make sure that the positioner has booted correctly. Run RTAC subsystem//PCHost//Gitane_ARCNET_COM diagnostics. If no error found then run COLLIMATOR subsystem//Collimator diagnostics. If no error found then it is not the root cause or it is intermittent error. If no error found then check the collimator cable is well plugged on the back of the positioner backplane (J300). Check the Positioner Power Supply. Also, a hot tube can cause Collimator errors, check the generator oil pump.

Applies to

Models
Innova
Product
X Innova 4100 X Do not use---------4100 systems 4100 systems
Software
Unity 3D M4 SP2 Unity 3D M4 SP2 Fwd Prod. DL Release : DL_UNITY_4.17.5 Ghost DL : 21.22.1 DL : DL_Unity_4.17.5 DL Atlas Core : 4.2-8_5 RTAC Atlas Core : 4.2-8_5 IDC : 5.26 Calypso : 4.37 Calypso Uptime : CLYUPT.UNITY_4.15 Positioner : RR POS LBY 1.4.4U_13 Jedi : P636RC1
Component
INX100; Positioner Mother Regulation Board, part 2335152-2 Positioner Mother Regulation Board. part 2335152-2. Always use the most appropriate documentation for your system. The following FRU reference was taken from Direction 5192425-2-2EN Rev#1 and is provided for your reference from the Innova 2100-IQ, 3100-IQ, 4100-IQ Elegance IQ-table DVD documentation. To view the most recent appropriate and detailed documentation, go to MY WORKSHOP Always use the most appropriate documentation for your system. The following FRU reference was taken from Direction 5192425-2-2EN Rev#1 and is provided for your reference from the Innova 2100-IQ, 3100-IQ, 4100-IQ Elegance IQ-table DVD documentation. To view the most recent appropriate and detailed documentation, go to MY WORKSHOP FRU reference 2335152-2
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues