GE Healthcare · MRExcite225229452239

Running diagnostics on the Driver Module failed indicating that the Multi-Coil Board #1 was not present. 11x to 14x upgr…

Running diagnostics on the Driver Module failed indicating that the Multi-Coil Board #1 was not present. 11x to 14x upgrade.

Symptom

Unable to use the A port when scanning with the HD TRknee PA coil 11.0 to 14.0 upgrade See long description System upgrade where everything worked fine accept coils plugged into the A port. In this case, they only had the one coil, but could not get it to scan. System reporting that the Driver Module was detecting an invalid Configuration. Source of Fault is Multi-Coil Board #1 not present.

Error message

2252294 The Driver Module Has Detected an Invalid Configuration. Source of Fault is Multi-Coil Board #1 Not Present.-> Dec 20 Thu 18:40:55 2252294 NSP SCP:DCB 304 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HD TRknee PA Scan: 52239/1 The Driver Module Has Detected an Invalid Configuration. Source of Fault is Multi-Coil Board #1 Not Present.-> See Long Description for Extended log... Error Code as viewed through the Service Browser with the Service Key installed: Dec 20 Thu 18:40:55 2252294 NSP SCP:DCB 304 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HD TRknee PA Scan: 52239/1 The Driver Module Has Detected an Invalid Configuration. Source of Fault is Multi-Coil Board #1 Not Present.-> Extemded Log viewed by clicking on the Hyperlinked Error code above: 2252294 ERROR: EM_DCB_CONFIG_FAIL The Driver Module Has Detected an Invalid Configuration. Source of Fault is %b. Extended Error: The requested Driver configuration has been determined to be invalid. The source of the fault will give specific information regarding the fault. Fault could be caused by hardware or software. Possible cause(s): Driver Modul

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
5141870-2 Cable Run 1203 HDx Systems
Area
MR

Full cause & step-by-step resolution