GE Healthcare · X-ray VascularInnova16000193000014200013200207
Single Shot Nosub mode, x-ray aborted.
Symptom
Single Shot Nosub mode, x-ray aborted. No fluoro. X-ray aborted by error 200013. Single Shot Nosub mode, x-ray aborted. No fluoro. X-ray aborted by error 200013.
Error message
Several foot switch short taps, errors 1600019, 3000014, 200013, 200207 Several foot switch short taps in a row, errors 1600019: An asynchronous error has been detected by the IP 3000014: XRSubSysMgr Software debug information 200013: XRAcqSync Time out while waiting for an event or an actor to respond. 200207: XRAcqSync Software Debug information See long description. Several foot switch short taps in a row, errors 1600019: An asynchronous error has been detected by the IP, warning generated 3000014: XRSubSysMgr Software debug information: Debug: [PerfMonitoringMgr::handlePerfMonitoringMsg] timer expired for action ACQUISITION_START_PREPARE 200013: XRAcqSync Time out while waiting for an event or an actor to respond. Debug: [XRAcqSync::_handleASTimeout] START_PREP phase waiting for actors: XRImProc 200207: XRAcqSync Software Debug information Debug: [XRAcqSync::_handleASTimeout] STOP_PREP phase waiting for actors: XRImProc
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- Unity 3D M4 SP2 Unity 3D M4 SP2 Fwd Prod. DL Release : DL_UNITY_4.17.5 Ghost DL : 21.22.1 DL : DL_Unity_4.17.5 DL Atlas Core : 4.2-8_5 RTAC Atlas Core : 4.2-8_5 IDC : 5.26 Calypso : 4.37 Calypso Uptime : CLYUPT.UNITY_4.15 Positioner : RR POS LBY 1.4.4U_13 Jedi : P636RC1
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.