GE Healthcare · X-ray VascularInnova105008210000146001051000025

The Omega table top locks do not release and the customer is unable to move the table top.

Symptom

The Omega table top locks do not release and the customer is unable to move the table top.

Error message

1050082 WR Positioner Power PIB 24V TBL POWER Debug: - PIB: 24V table PRD failed Omega 1050082 WR Positioner Power PIB 24V TBL POWER Debug: - PIB: 24V table PRD failed

error Code = 1000014, error Text = SID_SSD out of range, debug Text = SID out of range, recovery Class = IH, major Function /= XRImChainPos System impact: X-ray unavailable. Displayed message : "SID pb, use lift orange button to move detector to highest position". Note: This error is also logged when the positioner is resetted.

Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis. Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.