GE Healthcare · X-ray VascularInnova105008210000146001051000025
The Omega table top locks do not release and the customer is unable to move the table top.
Symptom
The Omega table top locks do not release and the customer is unable to move the table top.
Error message
1050082 WR Positioner Power PIB 24V TBL POWER Debug: - PIB: 24V table PRD failed Omega 1050082 WR Positioner Power PIB 24V TBL POWER Debug: - PIB: 24V table PRD failed error Code = 1000014, error Text = SID_SSD out of range, debug Text = SID out of range, recovery Class = IH, major Function /= XRImChainPos System impact: X-ray unavailable. Displayed message : "SID pb, use lift orange button to move detector to highest position". Note: This error is also logged when the positioner is resetted. Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis. Arcnet communication error issue. error Code = 600105, error Text = ARCNET Communication Failure followed by error Code = 1000025, error Text = ARCNET Communication Failure. Mainly due to Positionner, can be seen in application or in boot phasis.
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.