GE Healthcare · X-ray VascularInnova50213
Lift failure, dose is minimized, call service.
Symptom
Lift failure, dose is minimized, call service. Detector lift will not move up or down. Even by using the emergency backout button. Lift failure, dose is minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button. Lift failure, dose is not minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button.
Error message
error Code = 50213, error Text = Detector lift failure, debug Text += 4A77, recovery Class = WR, major Function /= XRImChainPos System impact: Lift lock-up. Displayed message : "Lift failure, dose is not minimized, call service asap".
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.