Gating does not work and the message The PAC does not respond Due to The PAC Fiber Optic Cable's Transmit and Receive Li…

Gating does not work and the message The PAC does not respond Due to The PAC Fiber Optic Cable's Transmit and Receive Lines Transposed (Connector Reversed).

Symptom

Gating does not work and the message The PAC does not respond displayed during system boot or TPS reset. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Power up Test and SPU-->PAC Loop back Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Powerup Test and SPU-->PAC Loopback Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay.

Error message

2254703 SPU TO PAC LOOPBACK FAILED. The SPU did not receive data back from the PAC during the SPU to PAC loopback test. Incorrect count of data received 0. Data count expected 8 --> Extended Error: The SPU did not receive data back from the PAC during the SPU to PAC loopback test. Incorrect count of data received. %d. Data count expected %d. Possible cause(s): PAC

2254702 PAC DIAGNOSTIC TIMEOUT. The PAC did not return a result to the SPU for the last PAC diagnostic test in the allotted 8 second time. Try running PAC datapath tests or other PAC diagnostics. Extended Error: Verify that the PAC module has power. Run SPU-->STIF and SPU-->PAC loopback tests to see if the MGD chassis can talk to the PAC module. Check fiber optic connections between the STIF and the PAC. Possible cause(s): PAC

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
46-328079G9 FO (Fiber Optic) CABLE ASSY (711M/712M) PP1-J91 TO MG2-A11-J9,10 & MG2-A33-J11,12,13 46-328079G9 Fiber Optic Cable Assembly: (711M/712M) PP1-J91 TO MG2-A11-J9,10 & MG2-A33-J11,12,13 46-328079G9
Area
MR