Excite System fails to scan.

Symptom

Unable to Scan With The Download of a Prescribed Series Failed Message

Error message

2248102 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset. See Long Description 2248102 qstmr1 NSP APS:processPkt 896 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset.

2220023 Coil 8HRBRAIN has 8 Channels, Product Supports only 6 Channels, Coil Not Compatible.

2246325 DRF1 DSP0 in Slot 7 failed the DPR Semaphore Test: Failure code: 17 (Cannot get semaphore from DSP side) see long description When running MGD diags see the following error: 1086244249 0 1 Thu Jun 3 06:30:49 2004 2246325 2 qstmr1 diagCoord drf_dpr_diags.c 227 Diagnostic: DRF 1 BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d2d8 Time: 30478 Object: APS Test Task DRF1 DSP0 in Slot 7 failed the DPR Semaphore Test: Failure code: 17 (Cannot get semaphore from DSP side) PCI Dynamic Base Addr: 0xc4100000 PCI Semaphore Addr: 0xdc0b0000 DSP Semaphore Addr: 0x2000000 Semaphore number: 0 PCI Semaphore Value: -1 DSP Semaphore Value: -505->

2246307 DRF1 HPI Dual Port RAM validation failure. A failure with Dual Port RAM was detected via the HPI interface. 1086244249 Thu Jun 3 06:30:49 2004 2246307 diagCoord DrfDiags.c 871 Diagnostic: DRF 1 BLD It

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.