GE Healthcare · MRExciteSigna22481021015122200232246325108624424930478
Unable to Scan With The Download of a Prescribed Series Failed Message
Excite System fails to scan.
Symptom
Unable to Scan With The Download of a Prescribed Series Failed Message
Error message
2248102 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset. See Long Description 2248102 qstmr1 NSP APS:processPkt 896 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset. 2220023 Coil 8HRBRAIN has 8 Channels, Product Supports only 6 Channels, Coil Not Compatible. 2246325 DRF1 DSP0 in Slot 7 failed the DPR Semaphore Test: Failure code: 17 (Cannot get semaphore from DSP side) see long description When running MGD diags see the following error: 1086244249 0 1 Thu Jun 3 06:30:49 2004 2246325 2 qstmr1 diagCoord drf_dpr_diags.c 227 Diagnostic: DRF 1 BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d2d8 Time: 30478 Object: APS Test Task DRF1 DSP0 in Slot 7 failed the DPR Semaphore Test: Failure code: 17 (Cannot get semaphore from DSP side) PCI Dynamic Base Addr: 0xc4100000 PCI Semaphore Addr: 0xdc0b0000 DSP Semaphore Addr: 0x2000000 Semaphore number: 0 PCI Semaphore Value: -1 DSP Semaphore Value: -505-> 2246307 DRF1 HPI Dual Port RAM validation failure. A failure with Dual Port RAM was detected via the HPI interface. 1086244249 Thu Jun 3 06:30:49 2004 2246307 diagCoord DrfDiags.c 871 Diagnostic: DRF 1 BLD It
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.