GE Healthcare · MRExcite2218650
Legacy connector pins were damaged preventing a good coil connection.
Symptom
4 Channel Coils Have Some Channels Without Signal - 8 Channel Coils Work OK Customer is trying to scan with some older 4 channel coil configurations such as USCTLTOP, TORSOPA, etc. However, whenever they do, they get some of the channels completely dead. 8 Channel Head and NV Arrays are fine which indicates the receivers and multi coil hardware is all OK.
Error message
2218650 Coil ID Warning Information returned from the coil indicates the coil does not contain a valid ID Fri Sep 7 10:15:12 2007 Error: 2218650 Host: SLMER-LX Process: scn File: SCP:SRI Line: 575 Coil ID Warning. Information returned from the coil indicates the coil does not contain a valid ID. An older coil that has not been upgraded with coil ID
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2296093-2 1.5T Integrated MultiCoil Assembly
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.