GE Healthcare · MRExciteSigna224813022185922248122221863022185012246324

PSD download errors on any 8 channel coil Due to DRF Board.

Symptom

PSD download errors on any 8 channel coil System is configured as a 8 channel system. PSD download errors on any 8 channel coil System is configured as a 8 channel system.

Error message

2248130 APS DC Setup DRF Channels Error. DRFs can only support -x channels 2248130 Host: UMIMR3 Process: NSP File: APS:DacqControl::processPkt Line: 513 APS DC setup DRF channels error DRFs can only support -1 channels Software error, please perform a TPS Reset

2218592 The MrMail send of real time filter to TPS timed out Thu Jun 3 05:32:17 2004 2218592 qstmr1 scn msaUtil.m 175 The MrMail send of real time filter to TPS timed out.

2248122 Number of Receiver Prescribed is Over the DRF Capability. 2248122 Host: UMIMR3 Process: NSP File: APS:DrfDriver::setupChannel Line: 20 Number of receiver prescribed is over the DRF capability. Maximum receivers supported at this system is 7 Please change prescription.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2246324 DRF Dual Port RAM (DPR) failure. DRF2 DSP1 in Slot 6 failed the DPR test Fri Oct 19 10:21:48 2007 Error: 2246324 Host: HMR5 Process: diagCoord File: drf_dpr_diags.c Line: 125 Diagnostic: DRF 2 BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d250 Time: 254822 Object: APS Test Task

224

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2215142 DRF board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.