Using the wrong Auto Install Floppy Disk 2304922.
Symptom
Unable to Load Solaris Operating System on AXI Computer Version 2.6 Rev.5/98 Solaris would stop loading after the load process accessed the Auto-install floppy. The floppy drive was cleaned and operating normally (access LED had activity). HDD was replaced which prompted the Load-From-Cold. CDROM drive was working because questions pertaining to time zone, networking, etc would appear and accept input. These questions are extracted from the Solaris CD which is mounted in the CD drive. Version 2.6 Rev.5/98 Solaris would stop loading after the load process accessed the Auto-install floppy. The floppy drive was cleaned and operating normally (access LED had activity). HDD was replaced which prompted the Load-From-Cold. CD-ROM drive was working because questions pertaining to time zone, networking, etc would appear and accept input. These questions are extracted from the Solaris CD which is mounted in the CD drive.
Error message
Warning Could not find matching rule in rules. ok This message appears in a pop-up window during the Solaris load at the point when the floppy drive is accessed. Full message as follows; "Searching for JumpStart directory… rules.ok found on floppy. Copying information to disk... Using rules.ok from floppy. Checking rules.ok file... Warning: Could not find matching rule in rules.ok Press the return key for an interactive Solaris install program.." "Searching for JumpStart directory… rules.ok found on floppy. Copying information to disk... Usingles.ok from floppy. Checking rules.ok file... Warning: Could not find matching rule in rules.ok Press the return key for an interactive Solaris install program.."
Applies to
- Product
- Profile 4
- Software
- 7.8x Profile4 or ProfileHSR only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.