GE Healthcare · MRExcite22236212248759
Will not scan with 2248759 RF Amplifier Response Timeout Due to SRFD2 RF Amplifier.
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
2223621 Internal Software Error. Can't callback on undefined NMRID:SCT 2223621 lx-mr scn mrmail_connect_mgr.c 239 Internal Software Error. Can't callback on undefined NMRID:SCT. 2248759 RF Amplifier Response Timeout RF MDS peripheral to RF Amp/RF Interface link failure. 2248759 NSP SCP: RF amplifier response timeout. RF MDS peripheral - RF Amp/RF Interface link failure. Verify that the RF Amplifier or RF Interface is powered on and that the RF MDS peripheral is connected to the RF Amplifier or RF Interface.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.