CWA ID86837. SGA Power Supply Errors May be Erroneous.

Symptom

High Incidence of Returned ACGD Power Supply With No Defect Found Subject: MR SMILe: High incidence of ACGD Power Supply No Defect Found power supplies. Effectivity: All Signa systems with ACGD Power Supplies (GEHC p/n T3504MH, 2334862-2, 2317092-2, 2284857-2, 2272460-2 ). Problem Description: The Repair Development Center (RDC) has seen a high rate of No Defect Found field return power supplies during repair. SMILe - May 4, 2004

Error message

2244585 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready 2244585 "The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. If ENABLE test point on the front SGA-PS J2 connector is HIGH, replace the SGA-PS. Else, check the GP to SGA-PS cable, or replace the GP."

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution