GE Healthcare · MRExcite2248662

MDS Failures Due to Reversed Fiber Optic Cables TX MDS and the RX MDS at the STIF.

Symptom

New Install Getting MDS Failures After Booting. New install, when system was coming up it would indicate an error of SCP auto config MDS failure

Error message

2248662 MDS link failure. The SCP failed to receive an MDS transmission echo. Tue May 25 10:38:44 2004 2248662 1 MROP NSP SCP:MdsAutoConfig 1443 MDS link failure. The SCP failed to receive an MDS transmission echo. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0000 Transmitted byte: 0x10

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR