The file listed in the message will not be present on any HFO3 system because the Body b0dither calibration, which creat…
The file listed in the message will not be present on any HFO3 system because the Body b0dither calibration, which creates this file, is not run on HFO3.
Symptom
Body EPI White Pixel test Message The file was not found /usr/g/bin delay.espb.xyz.dat.body using default values While running the Body EPI White Pixel test the system posts the following status message, which is confusing to the FE: "The file was not found: /usr/g/bin delay.espb.xyz.dat.body using default values". When the FE saw this message he became confused. The test still ran, but the FE was concerned there was something wrong with his system such as an incomplete EPI calibration. PQR 13138382. While running the Body EPI White Pixel test the system posts the following status message, which is confusing to the FE: "The file was not found: /usr/g/bin delay.espb.xyz.dat.body using default values". When the FE saw this message he became confused. The test still ran, but the FE was concerned there was something wrong with his system such as an incomplete EPI calibration. PQR 13138382.
Error message
The file was not found: /usr/g/bin delay.espb.xyz.dat.body using default values This message appears in the message box area, not the message log. Hence there is no error code. The file was not found: /usr/g/bin delay.espb.xyz.dat.body using default values
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.