CWA ID86168 Unable to Scan with Can Network Errors. 3T Excite HD 12.0. 2252232 SCP detected a Node Reset on the CAN netw…
CWA ID86168 Unable to Scan with Can Network Errors. 3T Excite HD 12.0. 2252232 SCP detected a Node Reset on the CAN network.
Symptom
Unable to Scan with CAN Network Errors See long description MR SMILe: March 17, 2005 Certain conditions can infrequently occur with devices on the CAN Link that will result in CAN Link errors and the loss of communication with these devices. Devices such as Driver Module, RF-DIF Board, GP3 Module, ASC (UPM) components, HOS, and Patient Handling Power Supply can be affected. CAN Link errors in the log and general troubleshooting oftentimes implicates one device on the CAN Link and results in it's replacement. Changing the part usually fixes the problem, but sometimes it does not. Another device on the CAN Link is later determined to be disrupting communications.
Error message
2244575 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault Error: 2244575 Host: WMR2 Process: NSP File: SCP:Grad Line: 2683 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault.-> 2252232 SCP detected a Node Reset on the CAN network 2252232 Host: EX3T Process: NSP File: SCP:CanLinkHWO Line: 617 SCP detected a Node Reset on the CAN network. Reset Node is Master RRF-DIF. Thu Aug 16 02:53:42 2007 Error: 2244591 Host: EX3T Process: NSP File: SCP:Grad Line: 3011 The GP is reporting a SGA Power Supply OverCurrent Fault.- : 2244573 The GP is reporting %c Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage 2244573 "The GP is reporting %c Axis SGA Internal Wiring Fault and\n\ Internal Power Supply Undervoltage. Perform a TPS Reset.\n\ Else, replace the SGA." 2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU t
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.