Pular para o conteúdo principal
GE Healthcare · MRSignaExcite22410792240410

Keyboard And Mouse Not Working Due To Cable Pulled Out Of Wim.

Symptom

Mouse and Keyboard Quit Working See Long Description

Error message

Keyboard Diagnostics Failed Power Up Test

2241079 Workspace Interface Module received a bad opcode from Workspace PC over serial link. See Long Description Error: 2241079 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_wim Line: 171 Workspace Interface Module received a bad opcode from Workspace PC over serial link. Opcode = 0x0. First, run diagnostics to test Workspace Interface Module to Workspace PC communication. If error persists, check cable connections, power cycle Workspace Interface Module and reboot Workspace PC.

2240410 - Host to Workspace PC Process Error. See Long Description Error: 2240410 Host: lx-mr Process: hop_server File: cpb_proto.c Line: 991 Host to Workspace PC Process Error. The Host received a unknown opcode from the Workspace PC. Verify Host to Workspace PC communication cables. Shutdown and boot host and Workspace PC.

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution