Pular para o conteúdo principal
GE Healthcare · MRExcite221868002077006459699769522186242218682

Intermittent ROTMI Calculation Errors Due to too broad of a detection for radial prescriptions.

Symptom

Error Occurred During Geometry Calculations Error occurred during Geometry Calculations. Re-prescribe slices/slab in Graphic Rx Screen.

Error message

2218680 Incorrect Rotmi row Error Mon Feb 12 10:09:49 2007 Error: 2218680 Host: lx-cc Process: scn File: RxManagerIF.m Line: 886 Slice:FreqMatError:Incorrect Rotmi row[2] = 0.020770 0.064596 0.997695

2218624 Failed to execute the next event. Error: 2218624 Host: lx-mr Process: scn File: StateMachine.c Line: 200 Failed to execute the next event.

2218682 Error occurred during Geometry Calculations Fri Jan 19 10:56:01 2007 Error: 2218682 Host: MR1 Process: scn File: RxManagerIF.m Line: 918 . Re-prescribe slices/slab in Graphic Rx Screen.

2218681 Rotmi Calculation Error found during Download Fri Jan 19 10:56:01 2007 Error: 2218681 Host: MR1 Process: scn File: RxManagerIF.m Line: 868 Rotmi Calculation Error found during Download

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues