GE Healthcare · MRExcite2227994
Insufficient TPS memory installed to support large data collection pulse sequences back-to-back.
Symptom
Unable to Scan with Not Enough or Insufficient TPS memory available Errors See Long Description Trying to set up for another scan and the system errors stating the reference scan stopped Insufficient TPS memory. 2218528 Host: lx-mr Process: scn File: MsaScanOpsScrn.m Line: 1505 Reference Scan has stopped. Insufficient TPS memory available.
Error message
2227994 The memory or other resources may be insufficient. Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wai Mon Feb 5 11:06:05 2007 Error: 2227994 Host: mr3t1 Process: NSP File: MH_MBOX_VRE:../scan_prep.cpp Line: 7812 Exception Class: Unknown Severity: Unknown Function: Recon: PSD: LAVA-XV Coil: 8US TORSOPA Scan: 5497/12 The memory or other resources may be insufficient. If the images of previous scans are still being reconstructed, please wait
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.