GE Healthcare · MRExcite2248674
Intermittent MDS Link Failures while Scanning.
Symptom
Intermittent MDS Link Failures while Scanning. Getting MDS link failures while scanning. Reports error 2241608 GIP VCC signal is out of range. Note that the value was mere millivolts out of spec, but enough to trip out scans. I measured the voltage at 4.81 volts, nominally low, but barely in spec. Spec is: Low: 4.70V High: 5.250V.
Error message
2248674 MDS link failure The MDS fiber optic link appears to be broken between the RF and System cabinets. Mon Feb 2 13:27:28 2004 2248674 1 MR-OC0 NSP SCP:../MdsLinkHWO.cpp 983 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.-> Mon Feb 2 13:27:28 2004 2248674 1 MR-OC0 NSP SCP:../MdsLinkHWO.cpp 983 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.