GE Healthcare · MRExcite10033

Cannot Archive Due to MOD drive.

Symptom

MOD Drive will not attach for archive. The system is a 5X Horizon system that was upgraded to use the Maxoptix MOD drive and media. An FE tried to order the Maxoptix T5-2600 drive thru Cares, but due to end of life issues, the drive changed up to a Sony MOD drive. After installing the new Sony drive, the system would recognize the drive as a Sony but could not attach or archive using the drive.

Error message

10033 There is no media in the drive Mon Apr 26 13:46:16 2004 Error: 10033 Host: MR-MN Process: arsjob File: ArchiveDevice.c Line: 4780 There is no media in the drive.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2318598 Sony MOD Drive Model SMO-F551-SD Sony MOD Drive 2318598 (Model #SMO-F551-SD) Sony MOD Drive 2318598 (Model #SMO-F551-SD)
Area
MR

Full cause & step-by-step resolution