Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova50108

All Gantry axis's will not drive

Symptom

INX100, Lost All Gantry axis's drive or NO movements The Gantry axis's will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor The Gantry axis' will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor

Error message

error Code = 50108, error Text = POS SMA: power voltage failure, debug Text += pos-error=4AB9, hw-status=0001, recovery Class = DO, major Function /= LC Gantry System impact: Degraded gantry motion. Displayed message : None. 0000 FRT_L 0001 FRT_PVT 0002 FRT_ARC 0003 FRT_II

CARC error Code = 50108, error Text = POS SMA: power voltage failure, debug Text += pos-error=4AB9, hw-status=0002, recovery Class = DO, major Function /= LC Gantry C_ARC Degraded Motion displayed. Gantry Motion may or may not be inhibited. Error code 50108 HW = 0002 FRT_ARC. System impact: Degraded gantry motion. Displayed message : None. 0000 FRT_L 0001 FRT_PVT 0002 FRT_ARC 0003 FRT_II

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Software
$, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
Area
X-ray Vascular