Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

600104 Collimator Control error Acquisition aborted; Collimator CAN communication failure 600104 Collimator Control error Acquisition aborted; re-start AR XRBeamForm control Component(s)_Collimator 1 63 Collimator CAN communication failure XRBeamForm.cpp 2068

Innova: 600104 4F76 Collimator Can Heartbeat Error Innova Console: 4F76 Xray Abort Reposition coll blades. Patient Dose Mismatch.. on the SUIF Error Log: 600104 Collimator Control error Acquisition aborted; re-start. AR. Collimator CAN communication failure. 600104 AR XRBeamForm Collimator Control error Acquisition aborted; re-start Debug: XRBeamForm CAN Heart Beat Failure See LD for complete error. Innova Console: 4F76 Xray Abort Reposition coll blades. Patient Dose Mismatch.. on the SUIF Error Log: 600104 Collimator Control error Acquisition aborted; re-start. AR. Collimator CAN communication failure. 600104 AR XRBeamForm Collimator Control error Acquisition aborted; re-start Debug: XRBeamForm CAN Heart Beat Failure Check the power, Have found the 24v going to the footswitch to cause this issue by loading down the power supply..

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues