Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
Symptom
Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76
Error message
DL 810006, error Text = Stop sequence timed out FOLLOWED BY 1610017, error Text = Unable to get property as requested, debug Text += Last frame Id FOLLOWED BY 1610006, error Text = Stop sequence timed out. Check 1600010 er System impact: exam failure. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service" Look at Cause error 1600010 System impact: exam failure. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". 810006 AR XRDataMgr Stop sequence timed out Debug: (null) 1610017 DO XRImProc Unable to get property as requested Debug: user StopSequence NOT done due to: Last frame Id 1610006 AR XRImProc Stop sequence timed out Debug: (null)
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- DL IMAGE DISKS DL IMAGE DISKS see long DL 36GB IMAGE DISKS 5144906 SCSI drive DL1 Image Disks Kit includes (2) 36GB image drives (ST336705LW), rack, fans, and cage. If no stock check part 2386702, also check SN4568B for alternate parts. P.Liebau 05/2009 DL 72GB IMAGE DISKS 5129897 *** OLC 2nd Opinion Part *** Kit contains (2) 72GB drives along with the mounting bracket, SCSI cable and terminator, for DL5+ and DL6 IP image disks. 1 FRU replaces both image disks. Contact your local OLE for approval. Replaced by new part 5248652 but that does not include the SCSI cable. P.Liebau 02/2009 5199031 See SN4844 for kit contents. Replaces 2332254 and 2386700 for DL2 to DL6. P.Liebau 11/2010 5248652 Kit contains (2) 72GB image disks and SCSI terminator, does NOT include SCSI cable which is part of DL internal cable set 5129896. P.Liebau 02/2009
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.