MOD will only attach if the media is in the drive on boot up.
MOD drive failure.
Symptom
MOD will only attach if the media is in the drive on boot up. Attach will fail on any other attempt see long description for more details on this issue The system is an LX system running 9.1 M4 software and a SONY SMO-F551 MOD drive. The system will attach a MOD media on boot up if media is in the drive. If the drive is empty on boot, attach will fail with "no media in the drive" error. Additionally if they have a media in the drive on boot and it successfully attaches, they can save fine, but when the Customer detaches that media and tries to attach another, the attach fails with "no media in the drive". They can reboot and the system will then attach the media that it just failed to attach, but again if they detach that media and try to attach another, it fails with "no media in the drive". This is very reproducible. Checked the configs for archive in ~sdc/app-defaults/archive directory. The SONY driver files are there and SCSI.fol looks good. The FE had already tried install.mod to update these config's with no success. lx-mr 1# su - sdc {sdc@lx-mr}[1] scsistat Device 0 1 Disk SGI IBM DNES-309170Y FW Rev: SA30 Device 1 3 CD-ROM TEAC CD-ROM CD-532S FW Rev: 1.0A Device 1 4 Optical SONY SMO-F551-SD FW Rev: 1.04 Device 2 1 Disk ANALOGIC DASM-LCAM-3M FW Rev: 1.5 Device 2 4 Comm CenData ST-1008S FW Rev: V7.2
Error message
10033 There is no media in the drive Mon Apr 26 13:46:16 2004 Error: 10033 Host: MR-MN Process: arsjob File: ArchiveDevice.c Line: 4780 There is no media in the drive.
Applies to
- Models
- Signa, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2318598 Sony MOD Drive Model SMO-F551-SD Sony MOD Drive 2318598 (Model #SMO-F551-SD) Sony MOD Drive 2318598 (Model #SMO-F551-SD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.