OrthoMRI Coil locking device is broken

Symptom

OrthoMRI The coil moves even after the locking mechanism is locked Inside the gantry the knee coil is locked by the coil lock mechanism, sometimes also if the coil lock mechanism is open the coil is not perfectly locked: it’s not possible to remove it but it’s possible to rotate it. In this case, it’s necessary to place one or two plastic disks under the coil. Inside the gantry the knee coil is locked by the coil lock mechanism, sometimes if the coil lock mechanism is open the coil is not perfectly locked: it’s not possible to remove it but it’s possible to rotate it. In this case, it’s necessary to place one or two plastic disks under the coil.

Applies to

Product
OrthoMRI Escan XQ EscanXQ and EscanNE
Software
OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
Component
OrthoMRI LNR9102029000 Coil lock mech You can also try LNR9102029300 which is the blue handled one if the LNR9102029000 is not in stock.
Area
MR

Full cause & step-by-step resolution