What survey meter can be used with HHS PRIMARY BARRIER TRANSMISSION test?The Fluke 451B (p/n 5270475) or Victoreen 470A…
What survey meter can be used with HHS PRIMARY BARRIER TRANSMISSION test?The Fluke 451B (p/n 5270475) or Victoreen 470A (p/n TC750787) are the only Survey meters to be used in conjunction with a Rad m…
Symptom
What survey meter can be used with HHS PRIMARY BARRIER TRANSMISSION test?The Fluke 451B (p/n 5270475) or Victoreen 470A (p/n TC750787) are the only Survey meters to be used in conjunction with a Rad meter for Innova Primary Barrier HHS tests. The Ludlum meter (GE tool pool p/n 2381102) Survey meter is not approved or validated for use on our Innova (or AdvantxE) systems. The Fluke 451B (p/n 5270475) or Victoreen 470A (p/n TC750787) are the only Survey meters to be used in conjunction with a Rad meter for Innova Primary Barrier HHS tests. Questions have arisen re: the use of a Ludlum 2401-EW Survey meter for the HHS Primary Barrier tests. This meter is not approved or validated for use on our Innova (or AdvantxE) systems. The Ludlum meter (GE tool pool p/n 2381102) is used by our PET FE's. It utilizes a thin end-window GM (Geiger-Mueller) tube detector, which is not designed for our type of radiation testing.
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor