GE Healthcare · MRExcite224485522445852244586
2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready.
Excite HDx. 2244855 errorcode on Post 14M5 is the same error as 2244585 on 11X, G3 and 2244586 on 12X, E2, 14M4 (HDx, HDe).
Error message
2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. Same as 2244585 and 2244586. 2244855 errorcode on Post 14M5 is the same error as 2244585 on 11X, G3 and 2244586 on 12X, E2, 14M4 (HDx, HDe). 2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready. 2244855 errorcode on Post 14M5 is the same error as 2244585 on 11X, G3 and 2244586 on 12X, E2, 14M4 (HDx, HDe).2244855 The SGA Power Supply failed to go to Ready when commanded, or came out of Ready.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.