GE Healthcare · X-ray R&FAdvantx2133821
Tomolink - No Exposure Troubleshoot the unit and found the modification to the Advantx rad transition panel wa…
Tomolink - No Exposure
Symptom
Tomolink - No Exposure Troubleshoot the unit and found the modification to the Advantx rad transition panel was mis-wired per pg 13-2 of direction 2133821-100 tomo-link sevice manual. The unit would prep with the exposure switch and not the rotor switch (prep switch). This timed out the tomo start sequence.