ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Repl…
ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
Symptom
XRd, ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; calibration was not allowed to continue. See long description: The Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; calibration was not allowed to continue. The same error code was displayed in Applications when an exposure on that detector was attempted. The error code indicated a problem in the handshake between the IDC and the Advantx-E Titan Kernel when the handswitch was pressed to the prep position so troubleshooting was focussed on these two subsystems. The IDC software was reloaded and then the IDC was replaced with no change in the symptoms. The Titan Kernel software was reloaded with no change in the symptoms; all nodes booted up OK. A telnet session to the IDC during a bad pixel calibration attempt captured the following: XR_Entrance_Dose::setMaxDose(10000,0) XR_Entrance_Dose::setKvPeak(80,0) XR_Entrance_Dose::setFiltration(10,0) XR_Entrance_Dose::setGridPosition(0,0) XR_Entrance_Dose::setMaxDose(0,1) XR_Entrance_Dose::setKvPeak(0,1) XR_Entrance_Dose::setFiltration(10,1) XR_Entrance_Dose::setGridPosition(1,1) ADX_IF:: PREP STARTS - EXECUTION TIME of SEQUENCE TIMING : 0 -- sending to TITAN prepResp (2) --> PREP_NOK ADX_IF:: PREP ENDS - EXECUTION TIME of SEQUENCE TIMING : 0 ADX_IF:: STOP STARTS - EXECUTION TIME of SEQUENCE TIMING : 4950 IDC-ACQUISITION(1):IDC_IDLE_STATE XR_Entrance_Dose.uponStopPrep Stop Sequence IDC-ACQUISITION(1):IDC_IDLE_STATE Apollo.uponStopPrep Image 0 already released. Image 1 already released. Image 2 already released. Image 3 already released. ADX_IF:: STOP ENDS - EXECUTION TIME of SEQUENCE TIMING : 4950
Error message
6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply)
Applies to
- Models
- Revolution, Advantx
- Product
- Revolution XRd
- Software
- Gipeto Version 29.2
- Component
- ADS
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Revolution XR/d After a Load From Cold, the AWS exhibits strange operation and the SUIF does not have any Positioner Configuration and Calibration selections due to the use of the incorrect Jumpstart…
- Revolution XRd2X Worklist query on "All Systems in this Modality" does not retrieve anything after FMI 10791A is completed due to the default query set to DX.
- Revolution XR/d error 4280 when trying to make exposure to table or wallstand detector
- Revolution XRd 2X The HP Workstation reboots itself.
- ADXE RFX/SFX Legacy popping noises during angulation SN 3715
- Advantx Rad Compax 40E system with SG100 Tilting Vertical Bucky Stand.