The Infiniband communication between the MGD and the ICN was broken.

Symptom

Error message during VRE Configuration and Infiniband Board Port 1 LED Not Lit When FE performed VRE configuration, it returned the following message. "Performing configuration ICN2(MAC 00:30:48:DE:6A) WITHOUT IB Cable" But it completed and he can scan without any errors. He confirmed that the GREEN LED on the Infiniband board was not lit at port 1. He swapped the cable from port 1 to port 2. But it was the same. PQR 13089664 SPR MRIhc19969 When FE performed VRE configuration, it returned the following message. "Performing configuration ICN2(MAC 00:30:48:DE:6A) WITHOUT IB Cable" But it completed and he can scan without any errors. He confirmed that the GREEN LED on the Infiniband board was not lit at port 1. He swapped the cable from port 1 to port 2. But it was the same. PQR 13089664 SPR MRIhc19969

Error message

Performing configuration ICN2(MAC 00:30:48:DE:6A) WITHOUT IB Cable Message returned during VRE configuration procedure: "Performing configuration ICN2(MAC 00:30:48:DE:6A) WITHOUT IB Cable"

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.