Pular para o conteúdo principal
GE Healthcare · MRExcite2218527

Unable to start post contrast axial VIBRANT scan after sagittal VIBRANT multi-phase series due to lack of data acquisiti…

Unable to start post contrast axial VIBRANT scan after sagittal VIBRANT multi-phase series due to lack of data acquisition memory.

Symptom

Unable to start post contrast axial VIBRANT scan after sagittal VIBRANT multi-phase series Customer wanted to do an axial 3D post contrast scan after multi-phase vibrant, but the axial series could not be prepped due to not having enough memory, since the multi-phase vibrant was still in progress. The operator sees a message like "Not enough Recon Memory". The complaint wasn't really that multi-phase vibrant took too long, it really was that they couldn't prep the second series to collect the post contrast data at the right time. PQR 13090559.

Error message

2218527 Scanning has stopped. not enough memory available to start acquisition, please try again. 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again. Error: 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again.

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.