Pular para o conteúdo principal
GE Healthcare · MRExcite225236070727

Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2252360 The Driver Module did not detect the T/R DD board, which is required Error: 2252360 Host: MR03 Process: diagCoord File: CanDriverHWO Line: 2721 Diagnostic: DCB Diag Iteration: 1 of 1 Action: 1 Task ID: 0x3156958 Time: 70727 Object: SCP T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset.-> Extended Error: 2252360 ERROR: EM_SCP_TRDD_NOT_PRESENT T/R DD Board Not Detected. The Driver Module did not detect the T/R DD board, which is required. Please make sure the board is present and perform a TPS reset. Possible cause(s): Driver Module

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution