Fuji CR9501 - Image plate jams due to being powered from warm start.

Symptom

Fuji CR9501 - Image plate jams due to being powered from warm start. A CQA was filed for this. The problem is that MIS 11251, tells you to connect power for the Fuji Chest stand to Cb14/15/16/17 in the RAD HALF cabinet, which is WARM START power. This causes IP jams if the Advantx console is cycled on-off-on while the 9501 is moving an IP. See CQA 986635. A workaround could be to connect to TS1-7 (120-A_COld) and TS1-8 (120-B-COLD), bypassing the warm start breakers.

Applies to

Models
Advantx
Component
FUJI
Area
X-ray R&F

Full cause & step-by-step resolution