GE Healthcare · X-ray R&FAdvantx11251986635
Fuji CR9501 - Image plate jams due to being powered from warm start.
Symptom
Fuji CR9501 - Image plate jams due to being powered from warm start. A CQA was filed for this. The problem is that MIS 11251, tells you to connect power for the Fuji Chest stand to Cb14/15/16/17 in the RAD HALF cabinet, which is WARM START power. This causes IP jams if the Advantx console is cycled on-off-on while the 9501 is moving an IP. See CQA 986635. A workaround could be to connect to TS1-7 (120-A_COld) and TS1-8 (120-B-COLD), bypassing the warm start breakers.
Applies to
- Models
- Advantx
- Component
- FUJI
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required