Will not scan with 2248759 RF Amplifier Response Timeout Due to SRFD2 RF Amplifier.
Symptom
Unknown SRFDII Error Code 2254050 with RF Amplifier fault 80, 81, 82, 83, 84, 85, 86, 87, 88, 89, and 90-96 On Excite systems running 11.0 software, we have seen error code 2254050 with various "RF Amplifier fault" codes such as 80, 81, 82, 83, 84, 85, 86, 87, 88, 89. However there is no other information in the error message and we can't find this error code in any troubleshooting document. What does error code 2254050 mean on Excite systems? PQR # 13099995. See Long Description for an example of this error message. On Excite systems running 11.0 software, we have seen error code 2254050 with various "RF Amplifier fault" codes such as 80, 81, 82, 83, 84, 85, 86, 87, 88, 89. However there is no other information in the error message and we can't find this error code in any troubleshooting document. What does error code 2254050 mean on Excite systems? PQR # 13099995 Tue Jan 9 16:08:43 2007 Error: 2254050 Host: lx-mr Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2146 RF Amplifier fault 80.
Error message
2248761 RF Amplifier Error. The RF amplifier reset watchdog engaged See Long Description 2248761 NSP SCP: RF Amplifier Error. The RF amplifier reset watchdog engaged. This indicates either a microprocessor fault condition exists or a cold start is in progress. 2254050 RF Amplifier fault 80. Error: 2254050 Host: lx-mr Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2146 RF Amplifier fault 80.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.