Pular para o conteúdo principal
GE Healthcare · CT26000244750043142720423040491770582

Lightspeed 1.X - Octane

Symptom

Multiple ring artifacts on axial scans - document in RFS - critical issue Multiple ring artifacts on axial scans

Error message

Error: 260002447 Questionable DAS Channel Detected. DAS Output View Data did NOT Pass Selftest Tue Apr 12 17:45:37 2005 Error: 260002447 Host: DAS Process: DCB Firmware File: dcvtst.c Line: 791 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50043/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:45:37:060 Log Series: 32 Questionable DAS Channel Detecte 260002447 Questionable DAS Channel Detected. DAS Output View Data did NOT Pass Selftest Tue Apr 12 17:45:37 2005 Error: 260002447 Host: DAS Process: DCB Firmware File: dcvtst.c Line: 791 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50043/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:45:37:060 Log Series: 32 Questionable DAS Channel Detected. DAS Output View Data did NOT Pass Selftest Converter Board Card: 17 Row: 1b Channel: 14. Channel Value Was: 0 Counts Expected: 1427204 to 2304049 Counts Average of All Channels Is: 1770582.

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2137958-4 SDAS Convertor Board (replaces 2137958-3) May use 2137958-3U converter Board if no current stock of the 2137958-4 See long description for details Part #: 2137958-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -39999.60 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CONVERTOR WITH BLACK BEUTY CONNECTOR FRU: Yes Status: Orderable Comments: OBSOLETE PART case 10788898 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 06-26-07 REPLACES P/N 2137958-3 THE NEW -4 IS COMPATIBLE TO THE -3. MS20544 ON 3/7/03. ** FE ASSIST ** May use the USED part 2137958-3U converter Board if no current stock of the 2137958-4. (9/10/2007 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.