GE Healthcare · CTLightSpeedVCT2300070042601327535004350018260132810230007007
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
White, wide band artifact in axial and arc artifact in helical with no error. See Long Description for Images. White, wide band artifact in axial and arc artifact in helical with no error
Error message
Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA EXCLUSION CRITERIA *. Error: 230007004 can be used for any row, channel and module that is failing. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA Examples: Sat Sep 17 08:08:02 2005 Error: 230007004 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 584 Exclusion Test - Row 33:1B, Channel 900, Module 57 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Tue Aug 29 05:24:31 2006 Error: 230007004 Host: ct99 Process: badchanmgr File: BadChannelMap.cxx Line: 667 Exclusion Test - Row 1:32A, Channel 418, Module 27 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Error: 260132753 Digital Interface Board(s) Encountered An Error During ASIC Linearization See Long Description for complete error Mon Oct 31 07:18:07 2005 Error: 260132753 Host: orp Process: orp_subsys File: Ifb.cxx Line: 504 Mon Oct 31 07:18:06 2005 Host : Das Ermes # : 260132753 Exception Class : Secondary Severity : Sec/Soft File : Ifb.cxx Line# : 504 Function : No System Function Reported Scan Type : Axial Scan: 50043/5/7 Digital Interface Board(s) Encountered An Error Duri
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5196401 FRDM for Center Modules for HDAS-64 see long description 5196401 HDAS FRDM Center Modules (Qty 57 used) Part #: 5196401 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: V-RES FIELD REPLACEABLE MODULE KIT - INNER SLOTS HALO CONVERTER CARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.