GE Healthcare · CTDiscovery2601327535004350018
Scanner Hardware Stopped Scan
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260132753 Digital Interface Board(s) Encountered An Error During ASIC Linearization See Long Description for complete error Mon Oct 31 07:18:07 2005 Error: 260132753 Host: orp Process: orp_subsys File: Ifb.cxx Line: 504 Mon Oct 31 07:18:06 2005 Host : Das Ermes # : 260132753 Exception Class : Secondary Severity : Sec/Soft File : Ifb.cxx Line# : 504 Function : No System Function Reported Scan Type : Axial Scan: 50043/5/7 Digital Interface Board(s) Encountered An Error During ASIC Linearization DIFB : 4 Chan : 1430 Stage : 2 This error format was introduced in 07MW18.4_SP2-26.V40_H_H64_G_GTL with SP2 release . We no-longer need the decoder since this error report the FRDM as A/D 41 . Wed Jan 7 08:40:45 2009 Error: 260132753 Host: orp Process: orp_subsys File: Ifb.cxx Line: 890 Wed Jan 7 08:40:44 2009 Host : Das Ermes # : 260132753 Exception Class : Abort Severity : Secondary File : Ifb.cxx Line# : 890 Function : No System Function Reported Scan Type : Axial Scan: 50018/1/3 Digital Interface Board(s) Encountered An Error During ASIC Linearization DIFB : 21 A/D Board: 41 Chan : 3369 Stage : 3 Coeff : 22 Other Examples: From software releases prior to 2009, the following error fo
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5206402 FRDM - Outer Modules for CT-750 HD systems. See long description for more information. 5206402 FRDM for Outer Modules for CT-750 HD systems This is the detector module for outer slots for Discovery CT750 HD systems. Part #: 5206402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT HD FIELD REPLACEABLE MODULE KIT - OUTER SLOTS FOR CONVERTER CARD FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 8/12/08.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required