GE Healthcare · MR20013

OrthoMRI Foot/Ankle cable coil is broken.

Symptom

OrthoMRI Cannot run a scout OrthoMRI Cannot run a scout.

Error message

OrthoMRI 20013 Incorrect Signal. Varicap2 Calibration The entire message reads: 20013 Incorrect signal. Varicap2 calibration, the amplitude of the signal during calibration isn't in a wished form, there is something not OK, check the .rec files. Do you want to proceed?

Applies to

Product
OrthoMRI Escan (Unix Digital) Two easy ways to tell if the Escan Unix system is analog or digital. First, the analog service password is coil and the digital password is probe. Second, the analog daily QA phantom is a bottle where as the digital daily QA phantom is a large spherical ball.
Software
OrthoMRI Unix 2.x
Component
OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. There is a PSDB solution available for deselecting the automatic coil recognition for ALL OrthoMRI systems. If need be, follow that resolution. Then the system can be used as normal once the operator manually selects which coil is being used while on the patient set up page. The most common failure is the particular coil's cable. If multiple coils have this issue the fix could be the mechanical swing arm.