OrthoMRI Shoulder coil 1 will not tune without errors.
Symptom
OrthoMRI Getting an error when running a scout.
Error message
OrthoMRI 20013 Incorrect Signal. Varicap2 Calibration The entire message reads: 20013 Incorrect signal. Varicap2 calibration, the amplitude of the signal during calibration isn't in a wished form, there is something not OK, check the .rec files. Do you want to proceed?
Applies to
- Product
- OrthoMRI Escan XQ EscanXQ and EscanNE
- Software
- OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
- Component
- OrthoMRI LNR9101819001 TUNABLE SHOULDER COIL AND CAP KIT TUNEABLE SHOULDER COIL USED ON MR AND CAP KIT ROUTED TO USAI FOR CONSUMABLE COLLECTION PER MIKE DIGGINS 1-03, A. MASCH DO NOT DISCOUNT - PER SHAKUN. Added per E-Scan RSPL. Tune the coil to the magnet frequency using the capacitor kit. Then, calibrate the varicap value and RF180 pulse and save these in manual coil tuning. Check coil SNR.
- Area
- MR
Full cause & step-by-step resolution